Please follow these preparation guide lines.
• Understand your mounting requirements.
• Mount with lip seals down and away from the work area.
• Brackets should be kept as short and rigid as possible.
• Surfaces must be in good condition, clean, and free of dirt.
Remove paint from machined surfaces used for mounting.
• Alignment brackets must not be removed until
instructed.
• Machine travel can not exceed the encoder measuring length.
• Either limit machine travel or use correct length scale.
• Never mount with lip seals upward or towards work area.
3
Acu-Rite Companies Inc.
ENC 125 T/E
Mounting Preparation
“L” = Measuring length
+ 1.75” nominal
over travel
2.0
[50.8]
2.4
[61]
2.4
[61]
2.0
[50.8]
“L” = Measuring length
+ 1.75” nominal
over travel
ENC 125 T
ENC 125 E
Cable direction
• Determine the cable exit direction before installing the encoder.
• To change the cable exit direction; remove the cover plate and
rotate the cable 180°.
T-10 Torx screw (2)
Plug
Cover plate
Cable grommet