Det-Tronics UD10 FlexVu Explosion-Proof Universal Display Unit User Manual

Page 15

Advertising
background image

13

95-8661

4.2

4-20 mA lOOp CAlIBRATION
Both the input and output current loops of the UD10

are trimmed at the factory. They can also be trimmed

in the field for maximum accuracy using the following

procedures. If the detector connected to the UD10 is

HART enabled, its 4-20 mA output signal can also be

trimmed.

When the UD10 is used with a detector that supports

HART communication, the output of the detector should

be calibrated first.

HART Detector Signal Calibration
Navigate down the menu to Device Test > D/A (Digital

to Analog) Trim.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test
Device Test

Device Test

Self Test
Response Test
Loop Test
D/A Trim

D/A Trim

Zero Trim
Gain Trim

Select Zero Trim. When this screen is entered, a

warning message is presented. Select ENTER to

continue. When the message “Connect Reference

Meter” is presented, install the current meter on the mA

line between the detector and UD10. Select ENTER

to continue. When the message “Set Input Current to

4mA?” is presented, select ENTER to begin the Zero

Trim function. The detector will now set its 4 mA output

value. If the value indicated on the current meter is not

4.00 mA, enter the measured value into the UD10 using

the Previous and Next switches. The UD10 calculates

and corrects for the difference between the actual and

entered values. When the current meter value is at the

desired 4.00 mA, select ENTER to accept the new zero

trim value.

Select Gain Trim. Follow the same procedure for gain/

span calibration.

UD10 Input Trim
When the UD10 is used with a detector that supports

HART communication, an automated process can be

used to trim the UD10 input. Navigate down the menu

to “Input Loop Cal”.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test
Device Test

Display Setup

Alarm Setting
Mode Select
HART Option
RTC
RS485
Input Loop Cal

Upon entering Input Loop Cal, the UD10 commands

the detector to output 4 mA, and then automatically

calibrates its own input. The UD10 then commands the

detector to output 20 mA and subsequently calibrates

its own input.

If a non-HART detector is being used, the Input Loop

Cal may be performed with a mA current source or loop

calibrator connected to the UD10 Sensor Connector.

Follow the loop calibration instructions shown by the

UD10 for this procedure.

Advertising