K factor, Menu structure – Det-Tronics UD10 FlexVu Explosion-Proof Universal Display Unit User Manual

Page 67

Advertising
background image

4.2

95-8661

M-4

Determining Sensor life Remaining
At the time of calibration, the UD10 logs the sensor

mV signal. This value can be used for determining the

approximate sensor life remaining.

To view all the recorded mV signal values for the sensor,

from the Main Menu, navigate to:

Device Status > Calibration Log > Span.

To view the mV signal from the most recent calibration

only, from the Main Menu, navigate to:

Device Status > Device Info > Response.

A typical new sensor reads between 45 and 55 mV.

• At 21-55 mV, “Cal OK” is recorded in the Cal Log,

along with the Span value.

• At 15-20 mV, “Cal OK” is recorded in the Cal Log,

along with the Span value. In addition, “Weak

Sensor” is shown on the UD10 display for about

20 seconds. After 20 seconds, the “Weak Sensor”

message is no longer seen, but a “Weak Sensor”

status is recorded (Device Status > Fault/Status >

Status).

• At 14 mV or less, “Weak Sensor” is shown on the

UD10 display for about 20 seconds, then a Cal Fault

is shown. The Cal Log shows “Cal Fail” with a Span

value of 0.00 mV.

K fActor

If the system will be detecting a gas/vapor other

than the gas used in the actual calibration process,

a conversion K-Factor must be used. The K-Factor

can be entered prior to calibration by navigating to the

“Device Option” menu, then selecting “K-Factor”. Enter

the desired K-Factor and activate the “Enter” button.

The new K-Factor will be applied when the sensor is

calibrated.

Main Menu
Process Vars
Display Status
Device Status
Display Setup
Device Setup
Device Cal
Display Test

Device Setup

Device Option

Device Option

Gas Name
K-Factor

The UD10 communicates the K-Factor to the CGS

Interface Board during the calibration process, where

the proper correction is made to ensure accurate

calibration.

The actual effect of the K-Factor can be observed

as the span portion of the calibration is completed.

For example, suppose a K-Factor of 0.865 has been

programmed. When calibration is performed, the

UD10 will display 50% as span is accepted. It will then

apply the K-Factor, and the displayed value will change

to 43.3% LFL.

For additional information regarding K-Factors,

including a list of K-Factors for many common gases,

refer to Technical Bulletin number 76-1017.

menu structure

UD10 with CgS Series Sensor
Refer to the following menu when using the UD10’s

LCD display and internal magnetic switches.

When connecting a HART Communicator to the UD10’s

4-20 mA output, refer to the “UD10 HART” menu in

Appendix E.

menu HeLp

Status menus only allow the user to view the data.

The Setup menus allow the user to both view and

edit the data.

Advertising