Technical data

Pin assignment

Ready

Signal is LOW as long as the Protec P3000 is ready for measurements.

Calibration stable

Signal is LOW when the signal detected during a calibration had enough time to
stabilize and goes back to HIGH after the background signal had enough time to
stabilize

Leak

Signal is LOW if the preset trigger level is exceeded

Error

Signal is LOW if a warning or error message is active.

Caution

Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.

Open collector output

Active = low

Pin

Command (not editable)

4

Ready

5

Calibration stable

16

Leak

17

Error