5-2

Protec P3000 Messages

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Error
no.

Plain text message

Possible Reason

Recommendation

E1

24V of the MC50 too low

Fuse F1 on the motherboard has
blown

Replace fuse F1 on motherboard*

E2

24V of sensor heating too low

Fuse F2 on the motherboard has
blown

Replace fuse F2 on motherboard*

E3

24V II too low

Fuse F3 on the motherboard blown

Replace fuse F3 on motherboard*

W4

24V of the OPTION socket too
low

Fuse F4 on the motherboard has
blown

Replace fuse F4 on motherboard*

W5

5V of the sniffer is too low

Fuse F5 on the motherboard has
blown

Replace fuse F5 on motherboard*

W6

5V of the internal Leak too low

PRO-Check electronics defective

Replace PRO-Check reference leak or
call nearest INFICON service
representative!

E7

-15V of the MC50 is too low

Motherboard defective

Call nearest INFICON service
representative!

E8

15V of the MC50 is too low

Motherboard defective

Call nearest INFICON service
representative!

E9

High Voltage error

Problem in sensor electronics

Call nearest INFICON service
representative!

W11

Wise current unstable

Current from Wise Technology
sensor did not stabilize within
20 min. after start-up!
Protec P3000 possibly not being
used for serveral days, sensor
current should stabilize with more
run-time.

Restart the Protec P3000.
If problem persists, call nearest
INFICON service representative!

E12

Wise Sensor not ignited

Current from Wise Technology
sensor too low for more than 10
min after power on

Restart the Protec P3000, if problem
persists, call nearest INFICON service
representative!

E13

Problem in Heater Control

Heater control for Wise
Technology sensor defective.

Call nearest INFICON service
representative!

W14

Discharge gone out

Sensitivity problem of Wise
Technology sensor
or
No ambient helium present (e.g.
sniffer purged with nitrogen)

Warntime every 2h.

Caution

Make proof or calibration.

Restart with sniffer line connected to
fresh air

If problem persists, call nearest
INFICON service repesentative!

W15

Replace sniffer tip filter

Filter in the sniffer tip is clogged

Replace the filter in the sniffer tip!

W17

Fore pump service interval
expired!

> 10,000 operating hours since last
fore pump service

Replace membrane of diaphragm
pump!*

W18

Air filter service interval
expired!

> 10,000 operating hours since last
main air filter service

Clean or replace main air filter!

See Section

7.2

E19

No communication with Wise
ADC

Wise Technology sensor defective
or
CPU-board defective

Call nearest INFICON service
representative!