3-2

Operation of the Protec P3000

k

in

a

26

e Ch

ap

te

r

3

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Notice

If you have not purchased the PRO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to
Settings / Interfaces PRO-Check and set the PRO-Check to “disabled” to
prevent future warnings (see section

4.5.6

).

3.2

Controls on the main display unit

All set-up and control functions are integrated into the main display unit via the menu
structure. The functions of the 8 control keys are displayed on the LC display. During
measurements the main interface is the probe handle display giving all necessary
information for proper leak testing.

After start-up the Protec P3000 will automatically go into measurement mode.

Bar graph display

The currently detected leak rate will be display in a bar graph in logarithmic scale.
The currently selected trigger level is indicated by a black line, the currently selected
search level is indicated by a dotted line. If the search level is exceeded the shape
of a bell is displayed at the top of the display, if the trigger is exceeded the bell starts
to blink (to ring).

Fig. 3-2 Measurement screen

Pos. Description

Pos. Description

1

Audio volume buttons

4

Info button

2

Menu button

5

List of gas parameters buttons

3

Calibration button