Settings Sub-menu

In the settings sub-menu several settings which are important for the leak testing
procedure can be edited.

4.4.1

Vacuum & Access

In the Vacuum Access sub-menu setting monitoring functions like the

ZERO

function,

the flow limits and the contamination limit setting can be edited.

Zero time

The

ZERO

function is automatically updated when the background decreases so that

negative values will never be displayed, thus avoiding unintended missing of leaks.
The

ZERO

TIME

is the time the leak rate has to be negative in order for the Protec

P3000 to update the background level. The

ZERO

TIME

can be set to values between

1.0 and 9.9 s.

Default value: 5.0 s

Contamination Limit

The Protec P3000 offers a protection mode against high helium concentrations
(contamination with helium).

This feature should be used to prevent the sensor from accelerated aging when
seeing large helium concentrations very frequently. The

CONTAMINATION

LIMIT

can be

set to levels of

LOW

,

NORMAL

AND

HIGH

. The contamination limit function will track the

accumulated amount of helium sniffed over the last couple of seconds. If this
accumulated amount of helium exceeds the preset level (

LOW

,

NORMAL

,

HIGH

) the

message „contaminated“ will be displayed until the amount of helium in the system
has decreased again sufficiently. However, the lower the

CONTAMINATION

LIMIT

setting

the smaller the largest possibly detected leak will be.

Notice

You should start with the

CONTAMINATION

LIMIT

set to

LOW

. If you feel that

you get the „contaminated“ message too often, you may increase the

CONTAMINATION

LIMIT

by one level at a time (to get the best usage out of

your Wise Technology Sensor).

Default: N

ORMAL

The lower the contamination limit is set, the longer the life time of the Wise
Technology sensor of the Protec P3000.