Operation of the protec p3000 – INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 53
Advertising
Operation of the Protec P3000
3-19
k
in
a
26
e Ch
ap
te
r
3
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
9)
A message “Please wait…” will be displayed until the calibration is finished.
After completion of the external calibration a screen summarizing the results of the
calibration will be displayed. The old calibration factor as well as the new calibration
factor will be displayed. If warnings are active and have been acknowledged during
the calibration process, the message is endorsed with the information "with active
warnings".
Fig. 3-21 Request to sniff air during external calibration
Fig. 3-22 Results of external calibration