INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 60

Advertising
background image

4-5

Equipment Settings

k

in

a

26

e Ch

ap

te

r

4

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Editing gas equivalent
parameters

Back in the edit gas submenu scroll to the

HELIUM

PERCENTAGE

line item and press

EDIT

.

In the opening submenu the helium fill pressure, the fill pressure for the gas
equivalent (refrigerant) later on and the concentration of helium used can be entered.
In the lower right corner the

CORRECTION

FACTOR

(from helium to refrigerant) will be

displayed. If a set of parameters is entered that exceeds the boundaries of the Protec
P3000 the correction factor will be displayed in inverted colours. In this case adjust
the parameters until the correction factor is displayed in normal colours again. Press
OK if all parameters are as desired.

Notice

This menu should also be used when working with diluted Helium but still
displaying the leak rate as a Helium leak rate.

Notice

This menu should be used at any time when the Helium fill pressure in pre-
testing differs from the refrigerant fill pressure later on to display the correct
leak rate value, even if the leak rate is still displayed as a Helium leak rate.

Fig. 4-4 Selecting gases from the library

Fig. 4-5 Setting a refrigerant equivalent correction factor