4.4.4

Setting-up / editing an I•Guide Program

When selecting the

SETUP

PROGRAM

button a submenu page will open showing the

list of 10 pre-programmable I•Guide programs and their current status (enabled /
disabled). If a program is enabled it will be offered in the list of programs to select
from when operating in the I•Guide mode. To change the status of a program use the

ON

/

OFF

button on the left side of the display.

To edit the parameters of a program select the program by scrolling to the
corresponding menu line and open the selected program by pressing the “E

DIT

button.

In the opening

EDIT

PROGRAM

sub-menu the name of the program, the gas type

selected for this program, the number of locations to be leak tested, the minimum
measuring time for each location as well as the wait time between two leak tests
allowed for moving the sniffer to the next location is stated. In addition, the
summarized global trigger value corresponding to the allowable leak rate for the
complete unit under test is stored.

Each parameter can be edited by scrolling to the appropriate line item and selecting
the highlighted item by pressing the

EDIT

button.

Fig. 4-17 Setting-up the I•Guide Mode

Fig. 4-18 Editing an I•Guide program