18

KDS EZFlow 2040

Operation Settings and Procedures

(Continued)

6. DATE ON/OFF

1)

Press

to enter MODE SETTING

status.

2)

Press

or

to item No.5 DATE

ON/OFF.

3)

Press

into DATE ON/OFF menu.

(screen 14).

4)

Press

or

to turn the date

display which is on the main screen ON

or OFF.

5)

Press

to confirm selection and

return to MODE SETTING screen.

6)

Press

to return to main screen.

MENU

ENTER

SET

ENTER

SET

EXIT

CLEAR

Screen 14

Screen 15

7. ML OR DROP SETTING

1)

Press

to enter MODE SETTING

status.

2)

Press

or

to item No.6 ML OR

DROPS.

3)

Press

to enter INFUSION RATE

UNIT SETTINGS menu (screen 15).

4)

Press

or

to choose infusion rate

in ml/h or drop/min.

MENU

ENTER

SET

WARNING: Because the drop size of different fluid may vary, the infusion accuracy

under drop/min mode is not as accurate as ml/h mode. It is recommended to use

ml/h mode for most infusion applications to ensure the accuracy.

WARNING: Do NOT use the pump under drop/min mode in a mobile environment; the

drip-chamber sensor can’t count the drops properly in shaking or vibration status.

IMPORTANT: If drop/min is chosen for the infusion rate, the drip-chamber sensor MUST be

connected to the pump and clamped on the drip chamber. Note: the sensor MUST be

clamped above the liquid surface.

IMPORTANT: If drip-chamber sensor is not installed properly, infusion CANNOT start in

drop/min mode, DROP SENSOR NOT IN POSITION (screen 16) message will be dis-

played.