KDS EZFlow 2040

Screen 31

Alarm Messages

(Continued)

3. AIR IN LINE ALARM

EXIT

CLEAR

Screen 29

Screen 30

4. OCCLUSION IN LINE

If air bubble is detected in the I.V set during a

regular infusion, AIR IN LINE alarm will be

triggered with an audible and visual flashing

light. (screen 29) The pump will stop regular

infusion and switch to a KVO (Keep Vain

Open) infusion mode. Press

key to

silence the alarm buzzer. Infusion can be

resumed ONLY after the AIR IN LINE condi-

tion is cleared.

If occlusion occurs in the I.V set or in the nee-

dle area during a regular infusion, OCCLU-

SION IN LINE alarm will be triggered with

an audible alarm and visual flashing light.

(screen 30) Infusion can be resumed ONLY

when the OCCLUSION condition is cleared.

During a regular infusion, while there is no

medication left, an EMPTY ALARM will be

triggered to remind the user of an empty med-

ication reservoir. (screen 31)

5. EMPTY ALARM