Kds ezflow 2040, Bolus function – KD Scientific EZFLOW 2040 User Manual

Page 26

Advertising
background image

24

KDS EZFlow 2040

Bolus Function

PURGE

BOLUS

PURGE

BOLUS

Screen 24

IMPORTANT: There will be no interruption for current infusion during the BOLUS setup.

WARNING: If the Bolus volume exceeds the remaining volume to be infused on main setup

screen, the pump will display ‘BOLUS TOO MUCH!’ and return to the main screen. The cur-

rent bolus setup will be cancelled!

ENTER

SET

ENTER

SET

ENTER

SET

Screen 25

1)

During normal infusion, an additional

dose of medication can be infused into

patient by holding

button.

BOLUS dosage can be preset by holding

for 2 seconds to enter SETUP

BOLUS (screen 24) menu.

2)

Press

or

to set preferred BOLUS

RATE in ml/h then press

to

confirm the rate setting.

3)

Press

or

to set preferred BOLUS

VOLUME in ml. Minimum bolus volume

is1 ml; maximum bolus volume is

9999 ml. Press

to confirm the

volume setting.

4)

Pump will remind the user by asking

“BOLUS?”, (Screen 25).

5)

Press

to confirm bolus setting.