Kds ezflow 2040, Bolus function – KD Scientific EZFLOW 2040 User Manual

Page 27

Advertising
background image

25

KDS EZFlow 2040

Bolus Function

(Continued)

PURGE

BOLUS

Screen 26

PURGE

BOLUS

ENTER

SET

PURGE

BOLUS

5)

To start a bolus, press

Bolus

infusion will start at the preset rate and

vol-ume. Bolus screen will appear

(screen 26) to remind user of the deliv-

ered bolus volume. Once the delivered

bolus volume reaches the preset bolus

volume, the pump will automatically

switch back to the current infusion rate;

normal infusion will continue.

WARNING: BOLUS volume will be

added into TOTAL VOLUME DELIVERED.

The preset bolus infusion rate and vol-

ume will be kept in the memory of the

pump. If the user wants to restart the

same bolus infusion, please hold

for 2 sec-onds to enter SETUP BOLUS

screen, then press

twice, and

start the same bolus by pressing

once.