Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual

Page 21

Advertising
background image

19

KDS EZFlow 2040

Screen 16

IV SET

Drop Rate Range

Equivalent Flow Rate

Operation Settings and Procedures

(Continued)

7. ML OR DROP SETTING

(Continued)

Screen 17

Screen 18

5)

If drop/min is selected, press

to

enter the SELECT IV SET menu as shown

on Screen 17 and Screen18 for different

IV set setting.

6)

Please verify the IV set type on the IV

set’s packaging then press

or

to find the correct IV set setting.

6)

Press

to confirm selection and

return to MODE SETTING screen.

7)

Press

to return to main screen

ENTER

SET

ENTER

SET

EXIT

CLEAR

WARNING: Failure to select correct IV set will result inaccurate infused volume counting

under drop/min mode.

IMPORTANT: Due to the design of the IV set, the minimum flow rate and maximum drop

rate are varied for different IV set under drop/min mode, check the following conversion

chart:

10 Drops/ml:

1/200 D/min

6/1200ml/h

15 Drops/ml:

1/200 D/min

4/800ml/h

20 Drops/ml:

1/200 D/min

3/600ml/h

60 Drops/ml:

1/180 D/min

1/180ml/h

IMPORTANT: For 60 Drops / ml IV set, the maximum drop rate is 1580 D/min; the maxi-

mum flow rate is 1580ml/h.