Kds ezflow 2040, Alarm messages – KD Scientific EZFLOW 2040 User Manual

Page 32

Advertising
background image

30

KDS EZFlow 2040

Alarm Messages

(Continued)

7. LOW BATTERY ALARM

Screen 33

Screen 34

If battery power is low while pump is operat-

ed with internal battery power supply,

BATTERY LOW ALARM will be triggered

with audible alarm and visual flash light.

(screen 33)

WARNING: If BATTERY LOW ALARM Is triggered, user must connect pump into a

electrical outlet immediately to work under plug-in power and charge the internal

batteries. Failure to do that may cause the pump working abnormally and result

serious injury to the patient!

When the BATTERY LOW ALARM is trig-

gered for the first time, pump will continue

operating for a while depending on the cur-

rent infusion rate. User MUST connect the

pump to the electrical outlet immediately. If

pump is not connected into an electrical outlet

within 30 minutes, PUMP STOPS IN 3

MINS alarm will be displayed on the screen

(screen 34) with audible alarm and visual

flash light. Pump will stop in 3 minutes after

the 3-minutes alarm is triggered.