8. DRIVE MECHANISM ERROR

Screen 33

IMPORTANT: If the battery depleted alarm is displayed. The pump’s drive can’t

work properly. The DRIVE MECHANISM ERROR alarm will be triggered in this case.

However, the user can clear this alarm by connecting the pump to electrical outlet and make

sure the power indicator on the front of the pump is green. Wait for 30 seconds. Then press

ON/OFF switch on the back of the pump to turn on the pump again.

WARNING: If the DRIVE MECHANISM ERROR alarm still appears in the start up proce-

dures, pump will not finish the normal operation. Pump cannot be used anymore unless

inspected by authorized service personnel!

If a mechanism error is detected during an

infusion, DRIVE MECHANISM ERROR

alarm will be triggered. Pump will stop infu-

sion and display error message on the

screen. (screen 33) . Once this message is

displayed, the user should disconnect the IV

set with patient immediately. Then press

ON/OFF switch on the back of the pump to

turn off the pump completely.