Kds ezflow 2040, Operation settings and procedures – KD Scientific EZFLOW 2040 User Manual

Page 22

Advertising
background image

20

KDS EZFlow 2040

Operation Settings and Procedures

(Continued)

8. ADJ. I.V SET (ACCURACY CALIBRATION)

WARNING: THE ACCURACY OF THE PUMP HAS BEEN CAIBRATED TO THE IV SETS

PACKAGED WITH THIS PUMP. The default setting is 0%. Other infusion sets should be pre-

pared according to its accompanying instructions and user should adjust this IV pump’s accu-

racy calibration according to manufacturer’s recommendation.. Failure to adjust the accu-

racy for other IV sets may result in an inaccurate infusion.

MENU

ENTER

SET

Screen 19

1)

Press

to enter MODE SETTING

status.

2)

Press

to item No.7 ADJ. I.V. SET.

3)

Press

to enter ADJ. I.V SET menu

(Screen 19). There are two options for

accuracy calibration:

SELF-CORRECTION ( For drop/min mode ONLY) If the self-correction function is selected

to calibrate the accuracy automatically, the drip-chamber sensor MUST be connected to the

pump and clamped on the drip chamber. The SELF-CORRECTION ONLY works under

drop/min mode; it will adjust the drop rate automatically according to the preset drop rate.

The SELF-CORRECTION DOES NOT work under ml/hr mode, the pump will NOT adjust

the flow rate automatically.

IMPORTANT: If drip sensor was not installed properly, self-correction function will not start.