KD Scientific Legato 100 Series Syringe Pump User’s Manual

Publication 5617-006 Rev D

5. Screw down the syringe clamp knob (6) until the

syringe barrel clamp (10) is tight against the

syringe barrel. Do not overtighten, especially

when using glass syringes.

6. Secure the syringe flanges by pushing the retaining

brackets (4,5) tightly against the syringe flanges

using the bracket clamping knobs (7,8). The

syringe plunger should be secured by the pusher

block retaining bracket and the flange of the

syringe should be secured by the syringe holder

block retaining bracket.

7.

Tighten the mechanical stop collars (9) on the

pump to prevent excess pusher block travel.

NOTE:

The syringe barrel clamp (10) [single

only] may need to be flipped depending on

your syringe size. To do so, completely

unscrew and remove the syringe barrel

clamp knob from the syringe block. Then

remove lock nut (11). Flip the bracket and

reassemble.

SYRINGE LOADING

2

5

4

1

3

9

7

8

10

6

> 30 cc syringes

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: For infuse/withdraw models, make sure

the plunger is secured in the retaining bracket

and the barrel of the syringe is secured in the

retaining bracket.

The Legato 100 Series can accommodate up to two syringes.

Depending on the mechanism, any type of syringe, glass,

plastic or stainless steel can be used. The pump can

accommodate from 0.5 µl to 60 ml single syringe and 0.5 µl

to 10 ml dual syringe. The syringe barrel clamp should be

flipped for syringes smaller than 30 ml. For syringes larger

than 30 ml, the syringe clamp should be used as shown.

1.

Squeeze the release button (1) and move the

pusher block (2) to the approximate length of the

syringe.

2. Unscrew the bracket clamping knobs (7,8) to

loosen the retaining brackets (4,5) on the pusher

block (2) and the syringe holder block (3).

3. Unscrew the syringe barrel clamp (10) by

unscrewing the syringe barrel clamp knob (6).

4. Place the syringe on the syringe holder block (3).

Ensure the barrel flange is within the syringe holder’s

block’s retaining bracket (5) and the plunger flange is

in the pusher block’s retaining bracket (4).

< 30 cc syringes

11

!

CAUTION:

Be sure mechanical stop

collars are positioned properly to prevent

the plunger from bottoming out on glass

syringes.

NOTE:

Syringe brackets (4 & 5) are only

available on Legato 110, 111 & 180.

10

Dual Syringe

Clamp