KD Scientific Legato 100 Series User Manual

Page 69

Advertising
background image

69

Appendices

K

D

S

L

e

g

a

to

1

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 100 Series Syringe Pump User’s Manual

Publication 5617-006 Rev D

TROUBLESHOOTING

1. Infusion Accuracy

To ensure infusion accuracy always use new syringe(s)

and measure syringe bore diameter and enter actual

dimensions in millimeters (mm) using the Custom Syringe

entry option. Additionally, make sure that the guide rods

and lead screw are properly lubricated.

NOTE: It is recommended that you save

your programs periodically by uploading

them to a PC.

KDS Legato 100 Series Display Error

Messages

2. Motor Stalled

This indicates that the pusher block travel has been

impeded. This may be caused by the syringe plunger

hitting bottom, a kink in the tubing (occlusion), syringe

plunger binding or any situation requiring more force to

the head of the syringe than the pump is capable of

delivering. To restart the pump, remove the cause of

stalling and/or increase the force setting, then press

RUN. The pump will continue the interrupted procedure

where it stopped.

3. Out of Range

A value was entered or encountered in a pump program

that was beyond the pump’s limits.

4. Power Failure Notification

If power to the pump is interrupted during operation, an

alarm will sound to alert the user to the interruption.

When power is restored, an onscreen message is

displayed to indicate the interruption.

5. Unit Won’t Power Up

Verify the the Power Source Switch on the rear of the

pump is set to DC-IN if powering through power supply

or RS-485 if powering in a chain.

The KDS Legato 100 Series Pump will notify you of a power failure with a

message when power is restored

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

Advertising