Nominal Minimum/Maximum Flow Rates for Various Syringes.

(Actual Limits will vary depending on syringe manufacturer)

Syringe Size

Syringe ID

Minimum Rate

Units

Maximum Rate

Units

0.5 µl

0.103 mm

0.54000

pl/min

596.496

nl/min

1 µl

0.1457 mm

1.14000

pl/min

1.19358

µl/min

2 µl

0.206 mm

2.28000

pl/min

2.38598

µl/min

5 µl

0.343 mm

6.36000

pl/min

6.61487

µl/min

10 µl

0.485 mm

12.7200

pl/min

13.2256

µl/min

25 µl

0.729 mm

28.7400

pl/min

29.8805

µl/min

50 µl

1.030 mm

57.4200

pl/min

59.6496

µl/min

100 µl

1.457 mm

114.900

pl/min

119.350

µl/min

250 µl

2.304 mm

287.400

pl/min

298.468

µl/min

500 µl

3.256 mm

573.960

pl/min

596.076

µl/min

1 ml

4.608 mm

1.14960

nl/min

1.19387

ml/min

1 ml

4.699 mm

1.19550

nl/min

1.24149

ml/min

3 ml

8.585 mm

3.99042

nl/min

4.14394

ml/min

5 ml

11.989 mm

7.78230

nl/min

8.08163

ml/min

10 ml

14.427 mm

11.2692

nl/min

11.7027

ml/min

LEGATO 180 MIN/MAX FLOW RATES