KD Scientific Legato 100 Series Syringe Pump User’s Manual

Publication 5617-006 Rev D

LEGATO 100 SERIES PROGRAM

EXPORT AND IMPORT

The Legato 100 Series pump permits the daisy chaining

of multiple pumps, as well as integration with/control by

an external computer through HyperTerminal. To

facilitiate operation in these modes, the Legato 100

Series application software includes a variety of

commands designed to simplify the export/import of

Programs between the pump and external devices. This

section outlines the use of these commands.

TRANSFER A PROGRAM BET WEEN PUMPS

Follow these instructions to transfer a Program between

two KDS Legato 100 Series pumps.

1.

Connect two pumps together via an RS-485

(firewire type) cable and set the address on the first

pump to ‘00’ using the Set Pump Address

command in the Settings menu.

2. Set the address on the second pump to something

other than ‘00’ using the Set Pump Address

command in the Settings menu.

3. From the Program Main screen, choose Program

Select then choose the Program you would like

to export.

4. Choose the

File Options

button.

5. Choose the Pump button and enter the address for

the pump you would like to receive the Program.

6. Choose Export Program to Pump.

7.

The pump will display a confirmation message in

the Message Area on the screen when the export

is complete. Note that long Programs may take a

few seconds to export.

8. Before disconnecting the pumps, verify that the

Program was successfully exported by choosing

Program Select from the Program Main screen on

the destination pump, then browse the available

Programs to verify that the exported Program is

shown in the menu.

NOTE: Do not attempt to Export/Import

Programs while the pump is running.

NOTE: Do not change screens while

Importing/Exporting a program.

NOTE: For additional details on the

physical interfaces used to connect the KDS

Legato 100 Series to external devices,

refer to the External I/O Connections

section in the Appendices.

Export Program confirmation screen showing successful export

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE:

If both pumps have software versions 2.0.0

or higher, programs can be transferred via

RS-485

If both pumps have software versions 1.0.4

or lower, programs can be transferred via

RS-485

If one pump has software version 1.0.4 or

lower and the other pump has 2.0.0 or

higher, programs must first be uploaded to

the PC, and then downloaded to the

desired pump using Syringe Pump

Communicator program