KD Scientific Legato 100 Series Syringe Pump User’s Manual

Publication 5617-006 Rev D

TVOLUME

Sets or displays the target volume. Configuration Run

screen mode only.

Command format:

tvolume [{target volume} {volume units}]

Query response:

<lf>[##:]Target volume not set<cr>

<lf>[##]<prompt>

or:

<lf>[##:] # xl<cr>

<lf>[##]<prompt>

WVOLUME

Displays the withdrawn volume. Configuration Run

screen mode only.

Command format:

wvolume

Query response:

<lf>[##:]# xl<cr>

<lf>[##]<prompt>

Time commands

CITIME

Clears the infused time. Configuration Run screen mode

only.

Command format:

citime

CTIME

Clears both the infused and withdrawn times.

Configuration Run screen mode only.

Command format:

ctime

CTTIME

Clears the target time. Configuration Run screen mode

only.

Command format:

cttime

CWTIME

Clears the withdrawn time. Configuration Run screen

mode only.

Command format:

cwtime

ITIME

Displays the infused time. Configuration Run screen mode

only.

Command format:

itime

Query response:

<lf>[##:]# seconds<cr>

<lf>[##]<prompt>

or:

<lf>[##:]##:##:##<cr>

<lf>[##]<prompt>

TTIME

Sets or displays the target time. Configuration Run screen

mode only.

Command format:

ttime [{target time}]

Query response:

<lf>[##:]Target time not set<cr>

<lf>[##]<prompt>

or:

<lf>[##:]# seconds<cr>

<lf>[##]<prompt>

or:

<lf>[##:]##:##:##<cr>

<lf>[##]<prompt>

WTIME

Displays the withdrawn time. Configuration Run screen

mode only.

Command format:

wtime

Query response:

<lf>[##:]# seconds<cr>

<lf>[##]<prompt>

or:

<lf>[##:]##:##:##<cr>

<lf>[##]<prompt>