R/f measurement – RaySafe Solo User Manual

Page 10

Advertising
background image

R/f MeASUReMent

Using R/f, RAd, dose or dent version.

for best accuracy, center the selected sensor field (R/f low, R/f high, RAd, dent low or dent high) and
position the long axis of the sensor field perpendicular to the anode-cathode axis of the tube.

1. SENSOR MENU

R/F low
(R/F and Dose)

Sensor for conventional low dose rate measurements < 1 mGy/s
(7 R/min), normally after a phantom.

R/F high
(R/F and Dose)

Sensor for conventional high dose rate measurements > 1 mGy/s
(7 R/min), normally before a phantom.

RAD

Sensor for conventional high dose rate measurements > 0.1 mGy/s
(0.7 R/min), normally before a phantom.

DENT low

Sensor for low dose rate measurements < 1 mGy/s
(7 R/min), normally after a phantom or for applications with high
(> 10 mm Al or equivalent) inherent filtration.

DENT high

Sensor for high dose rate measurements > 1 mGy/s
(7 R/min), normally before a phantom.

press STEP to step between sensors and SELECT to select a sensor.

10

RaySafe Solo User Manual – R/f Measurement