Vertical detector placement, Kvp on amx, Display codes – RaySafe Solo User Manual

Page 38

Advertising
background image

VeRtICAl deteCtoR plACeMent

When measuring on a mammography machine with a horizontal beam, such as MultiCare Platinum, let
the pressure from the compression paddle fix the RaySafe Xi mammography detector in position. Put the
detector with the cable end pointing downwards and with the MAM-window centered in the compression
paddle biopsy window, as shown below.

kVp on AMX

When measuring on AMX machines with R/f high, RAd or dent high, keep a short distance to get kVp instead
of kV.

dISplAy CodeS

DISPLAy

MEANING

* * *

Measurement has started.

- - -

no value can be calculated.

< 1 ms

less than 1 ms valid samples.

LoSignal

the dose rate is too low to calculate a value (increase mA or decrease fdd).

HiSignal

the dose rate is too high to calculate a value (increase fdd or decrease mA).

LoRange

Calculated kV is less than specified range.

HiRange

Calculated kV has exceeded specified range.

38

RaySafe Solo User Manual – tips and tricks