tIpS And tRICKS

ARRAnGeMent of MeASUReMent

Make sure nothing is in the X-ray field when measuring with the RaySafe Solo.

the fact that the dose is proportional to

1/d

2

(where

d is the fdd – focus to detector distance) may be used

when positioning the sensor.

d

SenSoR poSItIon

the solid state sensors in the RaySafe Solo detector are positioned approximately 7 mm beneath the detector
surface. the position is marked with a line on the aluminium housing. this position is used as a calibration
reference point for dose.

ReSet SWItCh

If the software hangs up, by any reason, activate the system reset by pressing a thin object through the small
hole beside the RS-232 connector on the side of the base unit.

If the detector hangs up when the battery is down and does not respond even when a charger is connected,
reset as described above and charge for some minutes before turning the instrument on again.