RaySafe i2 System User Manual – Using the real time display

6

7

8

1

2

3

4

5

Figure 6. The online view elements

LEGENd

1 personal dose overview (see page 14) access (tap anywhere on the white bar)

2 Scale, green zone

3 Scale, yellow zone

4 Scale, red zone

5 Accumulated procedure dose

6 peak dose rate indication

7 displayed name

8 displayed symbol (color of the dosimeter)

the displayed name identifies the dosimeters. It may be truncated due to lack of space.

By tapping the white bar of a dosimeter, you will enter the personal dose overview of that dosimeter.

the dose rate scale is divided in three zones:

• green zone indicates < 0.2 mSv/h (< 0.02 rem/h).

• yellow zone indicates 0.2 – 2 mSv/h (0.02 – 0.2 rem/h). If you are exposed to radiation in the yellow zone

frequently, you should take actions to minimize the dose exposure.

• Red zone dose indications 2 – 20 mSv/h (0.2 – 2 rem/h) and should not occur often during normal

procedures. If you are exposed to radiation in the red zone frequently, you should take actions to minimize
the dose exposure.