35

RaySafe i2 System User Manual – technical data

ChARACTERISTIC

mEASURE

log memory 2

3600 dose Rate samples with one-second resolution, cyclically
overridden. dose Rate is only measured above a 40 µSv/h threshold
limit

time resolution

1 second

time accuracy

Maximum error 2 seconds/24 hrs

power supply

via the cradle when connected to a computer via USB

expected battery life

At least 4 years based on “normal use”, which is defined as the
operating conditions under normal use are 8 hours per day, 5 days a
week and 52 weeks per year at an ambient temperature of 20°C.

Radio frequency

868.3 MHz, 918.3 MHz, 927.9 MHz (depending on purchased
configuration)

CRAdle

Table 5. Cradle technical specifications

ChARACTERISTIC

mEASURE

weight

50 g

dimensions

64x61x59 mm (wxHxd)

Cable length

1.5 m

power

via USB

Communication with computer

USB 2.0