RaySafe i2 System User Manual – Maintenance

MAIntenAnCe

CAlIBRAtIng tHe ReAl tIMe dISplAy SCReen

If the touch screen does not respond correctly to user interaction, a touch screen re-calibration may be
needed.

follow the instructions below to perform a re-calibration:

1. tap and hold on the start up screen that appears during real time display startup until the setup view

appears.

2. follow the instruction on the bottom of the screen and the subsequent instructions in the next views.

SynCHRonIze doSe HIStoRy to A USB MeMoRy

follow the instructions below to synchronize the dose history that is stored in the real time display to a USB
memory:

1. Insert a USB memory at the back of the real time display.

2. press the button Synch to USB. the synchronization will start. this may take a while, depending on how

much data that is stored in the real time display. Reset the dose history stored in the real time display by
pressing the button Reset.

CleAnIng And dISInfeCtIon

Cleaning and disinfection of this product is required periodically. guidelines for each are given below.

WARNING

Always isolate the equipment from the mains electrical supply before cleaning or disinfecting
to prevent electric shocks.

CAUTION

Never allow water or other liquids to leak into the equipment as this may cause electrical
short-circuits or metal corrosion.

Cleaning and disinfection techniques for both the product and the room must comply with all applicable local
laws and regulations.

CleAnIng

enameled parts and aluminum surfaces should only be wiped clean with a damp cloth and a mild detergent,
and then rubbed down with a dry woolen cloth. never use corrosive cleaning agents, solvents, abrasive
detergents or abrasive polishes. If you are not sure about the properties of a cleaning agent, do not use it.