Technical data, Dosimeter radiology, Dosimeter memory – RaySafe i2 System User Manual

Page 31: Dosimeter radiology dosimeter memory

Advertising
background image

31

RaySafe i2 System User Manual – technical data

teCHnICAl dAtA

doSIMeteR RAdIology

ChARACTERISTIC

mEASURE

operational quantity*

H

p

(10)

Reproducibility

10% or 1 µSv, whatever is greatest

dose Rate Range

40 µSv/h – 300 mSv/h

energy dependence X-/γ-rays

± 20% within n40 – n100
± 30% within n100 – n120

Angular dependence

± 5% within ± 5°
± 30% within ± 50°
+200%/-100% within ± 90°

temperature dependence

± 5% within 20-26°C
± 25% within 15-35°C

Battery voltage dependence

± 2% from fully charged until low battery shutdown

Response time

less than 1 s above 100 µSv/h, less than 5 s otherwise

position on body**

on torso outside lead apron

*) H

p

(10): personal dose equivalent at a depth of 10 mm according to ISo 4037.

**) position on body: the H

p

(10) measurement is only valid for a position on the torso outside any lead apron or

other protection. (to estimate effective dose to a user, one must use other means to estimate things like the
environmental radiation situation and the effectiveness of protection.)

doSIMeteR MeMoRy

the dosimeter has two dose data memories:

• the accumulated dose memory, where the dosimeter stores accumulated dose values every hour for the

entire lifetime of the dosimeter.