Figure 12, Exchanging the sample loop – Metrohm 930 Compact IC Flex ChS User Manual
Page 34
Advertising
3.9 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
930 Compact IC Flex ChS (2.930.1200)
Figure 12
Exchanging the sample loop
1
Pressure screw
Fastened to Port 6.
2
Sample loop
3
Pressure screw
Fastened to Port 3.
Exchanging the sample loop
1 Removing the existing sample loop
■
Unscrew the pressure screws (6.2744.010) at Port 3 and Port 6.
■
Remove the sample loop.
2 Installing a new sample loop
■
Fasten one end of the sample loop to Port 3 using a PEEK pres-
sure screw (6.2744.010).
■
Use the second PEEK pressure screw (6.2744.010) to fasten the
other end of the sample loop to Port 6.