7 injection valve, 8 suppressor, 9 detector – Metrohm 930 Compact IC Flex ChS User Manual

Page 95: Injection valve, Suppressor, Detector

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

930 Compact IC Flex ChS (2.930.1200)

■■■■■■■■

87

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1 - 50 MPa (1 - 500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0 - 49 MPa (0 - 490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism becomes active 2 minutes after system
start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

7.7

Injection valve

Actuator switch-
ing time

typ. 100 ms

Maximum operat-
ing pressure

35 MPa (350 bar)

Material

PEEK

7.8

Suppressor

Resistance to sol-
vents

No restriction

Switching time

typ. 100 ms

7.9

Detector

You can find the technical specifications for the detector in the detector's
manual.