Figure 20, Inline filter – removing the filter – Metrohm 930 Compact IC Flex ChS User Manual

Page 75

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex ChS (2.930.1200)

■■■■■■■■

67

Removing the filter

Figure 20

Inline filter – removing the filter

1

Filter housing
Inline filter housing. Part of the accessories
(6.2821.120).

2

Filter screw
Screw for the inline filter. Part of the acces-
sories (6.2821.120).

3

Filter (6.2821.130)
Packaging contains 10 pieces.

1 Shutting off the flow

Switch off the high-pressure pump in the software.

2 Removing the inline filter

Unscrew the both pressure screws from the inline filter.

3 Unscrewing the filter screw

Use two adjustable wrenches (6.2621.000) to loosen the filter screw
(20-2) from the filter housing (20-1) and unscrew it by hand.

4 Removing the filter

Remove the old filter (20-3) using tweezers.

Advertising