6 servicing the high-pressure pump, Servicing the high-pressure pump, Figure 14 – Metrohm 930 Compact IC Flex ChS User Manual

Page 61: High-pressure pump – parts

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex ChS (2.930.1200)

■■■■■■■■

53

5.6

Servicing the high-pressure pump

Figure 14

High-pressure pump – parts

1

Pressure screw, short (6.2744.070)
Fastened to the outlet valve holder.

2

Outlet valve holder

3

Fastening screw

4

Pump head

5

Inlet valve holder

6

Pressure screw, short (6.2744.070)
Fastened to the inlet valve holder.

7

Eluent aspiration tubing connector
Consists of a coupling with a pressure
screw.

8

Pressure screw, short (6.2744.070)
Fastened to the pump output.

Maintenance interval

The following parts of the high-pressure pump have to be serviced at least
once per year:

Inlet valve (6.2824.170)

Outlet valve (6.2824.160)

Piston seal (6.2741.020)

Zirconium oxide piston (6.2824.070)

Maintenance tasks can also be carried out if the following problems occur:

Unstable baseline (pulsations, flow fluctuations)

Advertising