Metrohm 930 Compact IC Flex ChS User Manual
Page 54
3.17 Conditioning
■■■■■■■■■■■■■■■■■■■■■■
46
■■■■■■■■
930 Compact IC Flex ChS (2.930.1200)
■
Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.
3 Starting equilibration
■
Start the equilibration in MagIC Net: Workplace
▶ Run ▶ Equi-
libration
▶ Start HW.
■
Visually inspect whether all capillaries and their connections from
the high-pressure pump to the detector are leak-tight. If eluent is
leaking out anywhere, tighten the corresponding pressure screw
further, or loosen the pressure screw, check the end of the capil-
lary and shorten it using the capillary cutter if necessary and
retighten the pressure screw.
4 Conditioning the system
Rinse the system with eluent until the required stability of the base-
line is attained (normally 30 minutes).
The instrument is now ready for measuring samples.