6 Troubleshooting

930 Compact IC Flex ChS (2.930.1200)

■■■■■■■■

81

6 Troubleshooting

Problem

Cause

Remedy

Marked drop in
pressure.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 12).

The baseline has a
large amount of
noise.

The eluent path has a leak.

Check the eluent path and fix the leak.

High-pressure pump – Con-
taminated pump valves.

Clean pump valves (see Chapter 5.6, page 53).

The eluent path is blocked.

Check the eluent path and fix the blockage.

High-pressure pump –
Defective piston seals.

Replace the piston seals (see Chapter 5.6,
page 53)
.

The pulsation absorber is
not connected or is defec-
tive.

Connect (see Chapter 3.8, page 24) or replace
the pulsation absorber.

The baseline is drift-
ing.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.2, page 12).

The organic solvent in the
eluent is evaporating.

Check the eluent bottle cap (see Chapter
3.5, page 18)
.

Constantly stir the eluent.

The pressure in the
system markedly
increases.

The inline filter
(6.2821.120) is blocked.

Replace the filter (6.2821.130) .

Suppressor blocked.

Regenerate the suppressor (see Chapter
5.10.3.2, page 71)
.

Note: 6.2821.180 pump tubing connection
with filter must be used.

Guard column – Blocked.

Replace the guard column (see Chapter 3.15,
page 40)
.

Separation column –
Blocked.

Regenerate the separation column (see
Chapter 5.13.4, page 79)
.

Replace the separation column (see "Con-
necting the separation column", page 44)
.

Note: Samples should always be microfiltered .