3 connecting the rinsing solution – Metrohm 930 Compact IC Flex ChS User Manual
Page 41
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
930 Compact IC Flex ChS (2.930.1200)
■■■■■■■■
33
1 Fasten the capillary labeled regenerant to the outlet of the inline filter
(6.2821.120) using a pressure screw (6.2744.070).
2 Use a pressure screw (6.2744.070) to fasten one end of the PEEK
capillary (6.1803.030) to the inlet of the inline filter.
3 Use a pressure screw (6.2744.070) and an adapter (6.2744.080) to
fasten the other end of the PEEK capillary to Port 1 of the Dosino.
4 Connect the FEP tubing (6.1805.120) from the bottle with the regen-
eration solution to Port 2 of the Dosino.
3.10.2.3
Connecting the rinsing solution
There are 2 options for rinsing the suppressor:
■
Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as a rinsing solution.
■
Rinsing solution via Dosino
Prepare the rinsing solution in its own bottle and transport with a Dos-
ino.
The rinsing solution is connected to the rinsing solution capillary.