6.1

■■■■■■■■■■■■■■■■■■■■■■

82

■■■■■■■■

930 Compact IC Flex ChS (2.930.1200)

Problem

Cause

Remedy

Injection valve – Valve
blocked.

Have the valve cleaned (by a Metrohm service
technician).

The retention times
in the chromato-
gram have changed
unexpectedly.

Separation column –
Diminished separating effi-
ciency.

Regenerate the separation column (see
Chapter 5.13.4, page 79)
.

Replace the separation column (see "Con-
necting the separation column", page 44)
.

The eluent contains gas
bubbles.

Ensure that the connectors for the eluent
degasser are firmly connected .

Deaerate the high-pressure pump (see Fig-
ure , page 39)
.

High-pressure pump –
Defective.

Request Metrohm Service.

Peak areas are
lower than
expected.

Sample – There is a leak in
the sample path.

Find and fix the leak in the sample path.

Sample – The sample path
is blocked.

Find and fix the blockage in the sample path.

Sample – The sample loop
is not (completely) filled.

Adjust the transfer time for the sample (see
"Determining the transfer time", page 77)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean the contact surfaces of the column chip
with alcohol.

Column chip is defective.

1. Save column configuration in MagIC Net™.
2. Notify Metrohm Service.

Individual peaks are
greater than
expected.

Sample – Sample carry-
over from the previous
measurement.

Check the rinsing time (see "Checking the rins-
ing time", page 78)
.

Suppressor – Insuffi-
cient flow of regen-
eration solution or
rinsing solution.

There is a leak in the sys-
tem.

Check all connections.

Suppressor – Backpressure
too high.

Clean the suppressor (see Chapter 5.10.3.3,
page 72)
or replace parts
(see Chapter
5.10.3.4, page 74)
.

The background
conductivity is too
high.

The suppressor is not con-
nected.

Connect the suppressor (see Chapter 3.10,
page 27)
.