7.10 Power connection

■■■■■■■■■■■■■■■■■■■■■■

88

■■■■■■■■

930 Compact IC Flex ChS (2.930.1200)

7.10

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

Internal fuse 3.15 A

7.11

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini DIN 8-pin (female) (for Dosino, stirrer, remote lines, etc.)

Detector

1 15-pin high density DSUB (female)

Column recogni-
tion

for an intelligent column

Leak sensor

1 jack plug

Further connec-
tions

Auxiliary

1 15-pin DSUB (female)

Service

1 15-pin DSUB (female)

7.12

Safety specifications

This instrument fulfills the following electrical safety requirements:

CE marking in accordance with the EU directives:

2006/95/EC (Low Voltage Directive, LVD)

2004/108/EC (EMC Directive, EMC)

Federal Inspectorate for Heavy Current Installations ESTI (Accreditation
Number SCESp 033)

Safety mark for certification type 2 in accordance with NEV (type
testing with market monitoring, EMC conformity)