Metrohm 798 MPT Titrino User Manual

Page 104

Advertising
background image

3.2. Remote control commands

798 MPT Titrino

100

3.2.2.16. Mode.Parameter.TitrPara.Pause

0...999 999

Parameters for DET, MET, SET: Pause time in s. Is waited off after the dis-

pensing of the start volume.

3.2.2.17. Mode.Parameter.TitrPara.MeasInput

1, 2, diff.

Mode.Parameter.TitrPara.Ipol -127...1...+127

Mode.Parameter.TitrPara.Upol -1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest ON, OFF

Parameters for DET, MET, SET:

Selection of the measuring input; valid with measured quantities pH and U.

"diff." means differential amplifier, see page 159.

With Ipol, the inquiries for the polarization current in µA (Ipol) and

.PolElectrTest are valid.

With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in

steps of 10 mV.

Besides .PolElectrTest is valid.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to an active state (titration or conditioning).

3.2.2.18. Mode.Parameter.TitrPara.Temp -170.0...25.0...500.0

Parameters for DET, MET, SET: Titration temperature in °C. If a Pt100 or

Pt1000 is connected, the temperature is measured continuously and the pa-

rameter .Temp is updated.

The temperature is used for the temperature correction in pH measurements.

3.2.2.19. Mode.Parameter.StopCond.VStop.Type abs., rel., OFF

Mode.Parameter.StopCond.VStop.V 0...99.99...9999.99

Mode.Parameter.StopCond.VStop.Factor 0...±999 999

Parameters for DET, MET, SET: Stop volume.

If an absolute stop volume (abs.) has been selected, the volume in mL is

valid.

A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size

*

factor

The factor is valid.

OFF means that the criterion is not monitored.


3.2.2.20. Mode.Parameter.StopCond.MeasStop pH:

0...±20.00, OFF

U:

0...

±2000, OFF

I:

0...200.0, OFF

Mode.Parameter.StopCond.UnitMStop read

only

Parameters for DET and MET: Stop when a measured value is reached. Entry

as pH value, in mV (with U and Ipol) and in µA (with Upol). The appropriate

unit can be viewed with .UnitMStop.

OFF means that the criterion is not monitored.

3.2.2.21. Mode.Parameter.StopCond.EPStop 1...9, OFF

Parameters for DET and MET: Stop when a certain number of EP's has been

found.