Metrohm 798 MPT Titrino User Manual

Page 85

Advertising
background image

3.2 Remote control commands

798 MPT Titrino

81

*Parameter

Tree part "Parameters for SET"

.

SET1

Control parameters for EP1

│ ├

.EP Endpoint 1

depends on meas.quant.

3.2.2.28.

│ ├

.UnitEp

Unit of endpoint

read only

ditto

│ ├

.Dyn

Dynamics

depends on meas.quant. 3.2.2.29.

│ ├

.UnitDyn

Unit of dynamics

read only

ditto

│ ├

.MaxRate

Maximum dosing rate

0.01...10...150, max.

ditto

│ ├

.MinRate

Minimum dosing rate

0.01...25.0...9999 ditto

│ ├

.Stop

Titration stop

│ │ ├

.Type

Type of stop criterion

drift, time

3.2.2.30.

│ │ ├

.Drift

Stop drift

1...20...999 ditto

│ │ ├

.Time

Switch-off delay time

0...10...999, inf

ditto

│ │ ├

.StopT

Stop time

0...999 999, OFF ditto

.

SET2

Control parameters for EP2, as for EP1

.

TitrPara

Titration parameters

│ ├

.Direction

Titration direction

+, -, auto 3.2.2.31.

│ ├

.XPause

Waiting time before start volume

0...999 999

3.2.2.32.

│ ├

.Start V

Start volume

│ │ ├

.Type

Type of start volume

abs., rel., OFF 3.2.2.15.

│ │ ├

.V

Volume for absolute start volume

0...999.99 ditto

│ │ ├

.Factor

Factor for relative start volume

0...±999 999

ditto

│ │ ├

.Rate

Dispensing rate for start volume

0.01...150.0, max. ditto

│ ├

.Pause

Waiting time after start volume

0...999 999

3.2.2.16.

│ ├

.ExtrT Extraction

time

0...999 999

3.2.2.33.

│ ├

.MeasInput Measuring

input

1, 2, diff.

3.2.2.18.

│ ├

.Ipol

Polarization

current

0...1...±127 ditto

│ ├

.Upol

Polarization voltage

0...400...±1270 ditto

│ ├

.PolElectrTest

Test for polarized electrodes

ON, OFF ditto

│ ├

.Temp

Titration temperature

-170.0...25.0...500.0 3.2.2.19.

│ ├

.TDelta

Time interv. for meas.acquisition

1...2...999 999

3.2.2.34.

.

StopCond

Stop conditions

│ ├

.VStop

Stop volume

│ │ ├

.Type

Type of stop volume

abs., rel., OFF

3.2.2.20.

│ │ ├

.V

Volume for absolute stop volume

0...99.99...9999.99 ditto

│ │ ├

.Factor

Factor for relative stop volume

0...±999 999 ditto

│ ├

.FillRate Filling

rate

0.01...150.0,

max. 3.2.2.23.

.

Statistics

Statistics

│ ├

.Status

Status of statistics calculation

ON, OFF 3.2.2.24.

│ ├

.MeanN

No. of individual determinations

2...20 ditto

│ ├

.ResTab Result

table

│ │ ├

.Select

original,delete n,delete all

ditto

│ │ ├

.DelN

Deletion of individual results

1...20 ditto

.

Presel

Preselections

│ ├

.Cond

Conditioning

ON, OFF 3.2.2.35.

│ ├

.DriftDisp

Display of drift during cond.

ON, OFF

ditto

│ ├

.DCor

Drift correction

│ │ ├

.Type

Type of drift acquisition

auto, man., OFF ditto

│ │ ├

.Value

Drift value for manual drift corr.

0.0...99.9 ditto

│ ├

.IReq

Request of Id's after start

id1, id1&2, all, OFF 3.2.2.26.

│ ├

.SReq

Request of smpl size after start

value, unit, all, OFF ditto

│ ├

.LimSmplSize

Limits for sample size

3.2.2.27

│ │ ├

.Status

Status of limit control

ON, OFF ditto

│ │ ├

.LoLim Lower

limit

0.0...999 999

ditto

│ │ ├

.UpLim Upper

limit

0.0...999 999 ditto

│ ├

.ActPulse

Output of a pulse

first, all, cond., OFF 3.2.2.28.