Metrohm 798 MPT Titrino User Manual

Page 122

Advertising
background image

3.2. Remote control commands

798 MPT Titrino

118

.MeasPt.Z1"6.34" Measured value (SET, MEAS) of the MPList, format de-

pends on the measured quantity

.MeasPt.Z2"25.8" Temperature of the MPList in °C

.EP.X"1.234" Volume coordinate of the EP

.EP.Y"5.34" Measured value coordinate of the EP

3.2.2.106. Info.ActualInfo.Display.L1

up to 32 ASCII characters

Info.ActualInfo.Display.L8

up to 32 ASCII characters

Info.ActualInfo.Display.DelAll $G

Lines of the display. The display can be written to from the computer. Pro-

ceed as follows:

1. Lock the display, see 3.2.2.119.

2. Delete the whole display (.DelAll).

3. For writing onto the display, the standard character set will be used.

4. Unlock the display, see 3.2.2.119

5. Delete the whole display (.DelAll).

6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.78) to refresh the

display.

$Q sends the contents of the corresponding display line.

3.2.2.107. Info.ActualInfo.Comport.Number read

only

$Q sends the comport number of the Titrino where the PC is connected.

3.2.2.108. Info.Assembly.CycleTime read

only

Info.Assembly.ExV read

only

Inquiries regarding basic variables of the assembly: Cycle time in s, volume of

the active Exchange Unit in mL.

3.2.2.109. Assembly.Bur.Rates.Forward.Selected digital, analog

Assembly.Bur.Rates.Forward.Digital 0...150,

max.

Assembly.Bur.Rates.Reverse.Selected digital, analog

Assembly.Bur.Rates.Reverse.Digital 0...150,

max.

Expel and aspirating rate.

Digital or analog control. With digital control, the inputted value applies (in

mL/min). "max." means maximum possible rate with the Exchange Unit in

current use.

Analog means rate control with the analog potentiometer on Titrino.

3.2.2.110. Assembly.Bur.Fill

$G, $H, $C

$G starts the 'FILL' mode of the buret function.

3.2.2.111. Assembly.Bur.ModeDis

$G, $S, $H, $C

Assembly.Bur.ModeDis.Selected volume, time

Assembly.Bur.ModeDis.V 0.0001...0.1...9999

Assembly.Bur.ModeDis.Time 0.25...1...86400

Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF

Assembly.Bur.ModeDis.AutoFill ON, OFF

Advertising