3.2. Remote control commands

798 MPT Titrino

116

C24.Unit: Unit of the assigned value

"%"

C26.ActN: Number of single results

"3"

C26.Mean: Mean (decimal places as for the result itself)

"3.421"

C26.Std: Standard deviation (decimal places as for the result + 1) "0.0231"

C26.RelStd: Relative standard deviation (in %, 2 decimal places)

"0.14"

3.2.2.100. Info.ActualInfo.Inputs.Status read

only

Info.ActualInfo.Inputs.Change read

only

Info.ActualInfo.Inputs.Clear $G

Info.ActualInfo.Outputs.Status read

only

Info.ActualInfo.Outputs.Change read

only

Info.ActualInfo.Outputs.Clear $G

Status sends the current status of the I/O lines, Change sends the information

regarding whether a change in status of a line has taken place since the last

clearing, Clear clears the change information. For the output, there is a con-

version from binary to decimal, e.g.

| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |

Line No.

| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |

Output: 2

1

+ 2

3

= "10"

1 means ON or change; 0 means OFF or no change.

The lines are assigned as follows (see also pages 165ff):

Inputs: Outputs:

0

Start (pin 21)

0

Ready (pin 5)

1

Stop (pin 9)

1

Cond. ok (pin 18)

2

Enter (pin 22)

2

Titration (pin 4)

3

Clear (pin 10)

3

EOD (pin 17)

4

Smpl Ready (pin 23)

4

L4 in TIP (pin 3)

5

pin 11

5

Error (pin 16)

6

pin 24

6

Activate, L6 in TIP (pin 1)

7

pin 12

7

Pulse for recorder (pin 2)

8

not used

(pin 6)

9

not used

(pin 7)

10

not used

(pin 8)

11

not used

(pin 13)

12

Smpl size out

(pin 19)

13

Result out

(pin 20)

3.2.2.101. Info.ActualInfo.Assembly.CyclNo read

only

$Q sends the current cycle number of the voltage measurement cycle, e.g.

"127". From the cycle number and the cycle time (see 3.2.2.108), a time

frame can be set up.

The cycle number is set to 0 on switching on the instrument, on every start

and for QuickMeas. It is incremented as long as the instrument remains

switched on.

3.2.2.102. Info.ActualInfo.Assembly.Counter.V read

only

Info.ActualInfo.Assembly.Counter.Clear $G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the

volume counter is set to zero.