3 status messages – Metrohm 798 MPT Titrino User Manual

Page 74

Advertising
background image

3.1. General rules

798 MPT Titrino

70

3.1.3 Status messages

In order to have an efficient control by an external control device, it must also

be possible to query status conditions; they provide information on the status

of the Titrino. The trigger $D initiates output of the status. Status messages

consist of the global status, the detailed status and eventual error messages,

e.g. $S.Mode.SET;E26. The global status informs on the activity of the proc-

ess, while the detailed status conditions show the exact activity within the

process.

The following global status conditions are possible:

$G

Go:

The Titrino is executing the last command.

$H

Hold:

The Titrino has been held ($H, key <meas/hold> or by an error which effects

the hold status)

$C

Continue:

The Titrino has been restarted actively after hold

$R

Ready:

The Titrino has executed the last command and is ready

$S

Stop:

A process has been aborted in an "unnatural manner". e.g. stopped or aborted

because there was an error.

Detailed status conditions

Status conditions of the global $G:

$G .Mode.DET .Inac:

Instrument at the beginning or at the end of a titration.

.Req .Id1:

Instrument in the DET mode, requesting Id1 after titration start.

.Id2:

Instrument in the DET mode, requesting Id2 after titration start.

.Id3:

Instrument in the DET mode, requesting Id3 after titration start.

.Smpl:

Instrument in the DET mode, requesting sample size after titration start.

.Unit:

Instrument in the DET mode, requesting unit of sample size after titration

start.

.Start:

Instrument in the DET mode, processing the start conditions.

.Titr:

Instrument in the DET mode, titrating.

$G .Mode.MET...

As DET.

$G .Mode.SET .Inac:

Instrument at the beginning or at the end of a titration.

.Req

.Id1:

Instrument in the SET mode, requesting Id1 after start.

.Id2:

Instrument in the SET mode, requesting Id2 after start.

.Id3:

Instrument in the SET mode, requesting Id3 after start.

.Smpl:

Instrument in the SET mode, requesting sample size after start.

.Unit:

Instrument in the SET mode, requesting unit of sample size after start.

.Start:

Instrument in the SET mode, processing the start conditions.

.SET1:

Instrument in the SET mode, titrating to the first endpoint.

.SET2:

Instrument in the SET mode, titrating to the second endpoint.

.Cond.Ok:

Instrument in the SET, conditioning, endpoint reached (after the first

startup from the standby mode).

.Cond.Prog:

Instrument in the SET mode, conditioning, endpoint not reached

(Conditioning progressing).

$G .Mode.MEAS .Inac:

Instrument at the beginning or at the end of a titration.

.Req .Id1:

Instrument in the MEAS mode, requesting Id1 after start.

.Id2:

Instrument in the MEAS mode, requesting Id2 after start.

.Id3:

Instrument in the MEAS mode, requesting Id3 after start.

.Smpl:

Instrument in the MEAS mode, requesting sample size after start.

.Unit:

Instrument in the MEAS mode, requesting unit of sample size after start.

.Meas:

Instrument in the MEAS mode, measuring.

$G .Mode.CAL .Inac:

Instrument at the beginning or at the end of a calibration

.Req.Temp:

Instrument in the CAL mode, requesting calibration temperature.

.Meas.Temp:

Instrument in the CAL mode, measuring calibration temperature.

.Req.Buf1:

Instrument in the CAL mode, requesting pH of buffer 1.

.Meas.Buf1:

Instrument in the CAL mode, measures buffer 1.