Metrohm 798 MPT Titrino User Manual

Page 112

Advertising
background image

3.2. Remote control commands

798 MPT Titrino

108

Format the memory card. The action is carried out if "$G" is transmitted to the

corresponding node.

Do not use blank characters before and after name!

3.2.2.60. MemoryCard.FreeMem read

only

Free memory on the card. $Q sends number of free bytes, e.g. "4928".

3.2.2.61. MemoryCard.BatteryChange $G

MemoryCard.BatteryChange.Date YYYY-MM-DD

Date for changing battery. The date will be set with "$G".

3.2.2.62. MemoryCard.List.Card.1.Name read

only

MemoryCard.List.Card.1.Bytes read

only

for each directory

List of all directories on the memory card with the following information:

.Name:

Name of the directory

.Bytes:

Number of bytes used by the directory

3.2.2.63. MemoryCard.List.ActDir.1.Name read

only

MemoryCard.List.ActDir.1.Mode read

only

MemoryCard.List.ActDir.1.Quantity read

only

MemoryCard.List.ActDir.1.Bytes read

only

MemoryCard.List.ActDir.1.Checksum read

only

for each method

List of all methods of the current card directory with the following information:

.Name:

Name of the method

.Mode: Mode

.Quantity: Measured

quantity

.Bytes:

Number of bytes used by the method

.Checksum: Checksum of the method, see 3.2.2.94.

3.2.2.64. Config.Monitoring.Validation.Status ON,

OFF

Config.Monitoring.Validation.Interval 1...365...9999

Config.Monitoring.Validation.Counter

0...9999

Config.Monitoring.Validation.ClearCount $G

Monitoring of validation.

.Interval: Time interval in days for validation.

.Counter: Time counter in days since last validation.

.ClearCount: Clears the above counter.

3.2.2.65. Config.Monitoring.Calibration.Status ON,

OFF

Config.Monitoring.Calibration.MeasInput

1, 2, diff

Config.Monitoring.Calibration.Interval 1...7...9999

Config.Monitoring.Calibration.Counter

0...9999

Monitoring of pH calibration.

.MeasInput:Measuring input.

.Interval: Time interval in days for calibration from the last calibration date.