3.1 General rules

798 MPT Titrino

75

E130

Wrong sample. For SET with preset titration direction the first

measured value lies behind the endpoint.

Exit: The error message disappears on next startup.

E131

No EP set for SET.

Exit: The error message disappears on next startup.

E132

Silo empty and it has been started with open silo or empty silo has

been opened.

Exit: Send a silo entry.

E133

Silo full.

Exit: Send new command.

E134

No method. A method, which is required from the silo memory or in TIP, does

not exist.

Exit: The error message disappears on next startup.

E135

Check temp.sensor in MEAS T or with activated temperature monitoring.

Exit: Correct error or &m $S.

E136

Same buffer in CAL. Measured value of the second buffer differs less than 6

mV from the measured value of the first buffer.

Exit: Correct error or &m $S.

E137

XXX Bytes are missing so that the method, the silo line could not be stored or

not enough RAM for running TIP.

Exit: Send new command.

E155

No new silo result (C24 or C25).

Exit: The error message disappears on next start or on recalculation.

E157

No sequence defined in TIP.

Exit. The error message disappears on next start.

E158

A second TIP has been called up in TIP.

Exit: The error message disappears on next start.

E160

No new temporary variable.

Exit: The error message disappears on next start.

E161

Measurement range of the secondary measured value (temperature)

exceeded. The primary measured value (pH, U, Ipol, Upol) can also be

unstable.

Exit: Rectify error or &m $S.

E166

Save lines is "OFF" although a submethod of TIP includes an assignment to

C24 or C25.

Exit: The error message disappears on next start. Attention: The data of this

sample will not be stored.

E172

In TIP, a QuickMeas was started, without defining a measuring quantity.

Exit: The error message disappears on next start or &Mode.QuickMeas $S.

E177

Accessing to the memory card, the card was not (properly) inserted.

Exit. The error message disappears on next start.

E178

The date of changing the battery of the card is expired.

Exit. The error message disappears on next start.

E180

Memory card write-protected.

Exit: Send new command.

E181

Memory card not formatted.

Exit: Send new command.

E182

Memory card not accessible.

Exit: Send new command.

E183

A directory with the same name exists already on the memory card.

Exit: Send new command.

Advertising