Metrohm 798 MPT Titrino User Manual

Page 110

Advertising
background image

3.2. Remote control commands

798 MPT Titrino

106

.Stirrer:

Switching stirrer on/off.

3.2.2.49. Mode.Parameter.Presel.MeasMode

pH, U, Ipol, Upol, OFF

Mode.Parameter.Presel.MeasInput

1, 2, diff.

Mode.Parameter.Presel.Ipol 0...1...±127

Mode.Parameter.Presel.Upol 0...400...±1270

Mode.Parameter.Presel.PolElectrTest ON, OFF

Mode.Parameter.Presel.Temp -170....25.0...500.0

Parameters for TIP: Selection of the measured quantity for manual measure-

ments in the inactive state, see 3.2.2.2. Selection of the measuring input

(MeasInput) applies to measured quantities pH and U. "diff." means differen-

tial amplifier, see page 159. With Ipol the requests for the polarization current

in µA (Ipol) and .PolElectrTest apply. With Upol the request for the polarization

voltage in mV (Upol) applies. Entry in steps of 10 mV. .PolElectrTest also ap-

plies. If the test for polarized electrodes is switched on (ON), it will be per-

formed on the change from the inactive state to an active state. The tempera-

ture applies to pH measurements.

3.2.2.50. UserMeth.FreeMem read

only

Memory space, available for user methods or silo lines. $Q sends the number

of free bytes, e.g.

"4928".

3.2.2.51. UserMeth.Recall $G

UserMeth.Recall.Name

up to 8 ASCII characters

UserMeth.Store $G

UserMeth.Store.Name

up to 8 ASCII characters

UserMeth.Delete $G

UserMeth.Delete.Name

up to 8 ASCII characters

UserMeth.DelAll $G

Management of the internal method memory: Load, store and delete methods.

An action is performed if "$G" is sent to the corresponding node just after en-

tering the name.

Do not use blank characters before and after method name!

.DelAll: Deletes all methods in the user memory.

3.2.2.52. UserMeth.List.1.Name read

only

UserMeth.List.1.Mode read

only

UserMeth.List.1.Quantity read

only

UserMeth.List.1.Bytes read

only

UserMeth.List.1.Checksum read

only

for each method

List of the methods in the user method memory with the following character-

istics:

.Name:

Name of the method

.Mode: Mode

.Quantity: Measured

quantity

.Bytes:

Number of bytes of the user memory used by the method

.Checksum: Checksum of the method, see 3.2.2.94.