Figure 6, Mounting the air tube: correct - incorrect – Metrohm 895 Professional PVC Thermomat User Manual

Page 21

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

895 Professional PVC Thermomat

■■■■■■■■

13

Figure 6

Mounting the air tube: correct - incorrect

3.2.3

Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them
in the 895 Professional PVC Thermomat and establish the tubing connec-
tions (see Chapter 2.1, page 6) as follows:

1 Inserting the measuring vessel

Fill distilled water into the measuring vessel.

Place the measuring vessel cover onto the measuring vessel.

Insert the measuring vessel into the openings provided on the 895
Professional PVC Thermomat. While doing so, carefully guide the
connector plug into the electrode connector.

Connect the white silicone tubing to the M8/olive tubing adapter
of the measuring vessel cover.

2 Mounting the tubing for the air supply

Screw the FEP tubing 250 mm to the air supply connections of
the 895 Professional PVC Thermomat.

3 Inserting the reaction vessel

Fill the reaction vessel with the sample.

After the required reaction temperature has been reached, insert
the reaction vessel with the mounted reaction vessel cover in the
openings provided on the 895 Professional PVC Thermomat.

4 Establishing the tubing connections

Connect the white silicone tubing that is connected to the mea-
suring vessel cover to the tubing connector of the reaction vessel
cover.

Screw the FEP tubing 250 mm which is connected to the M8/olive
tubing adapter of the 895 Professional PVC Thermomat to the
M8/M6 thread adapter of the reaction vessel cover.