3 cleaning the accessories, Cleaning the accessories – Metrohm 895 Professional PVC Thermomat User Manual

Page 34

Advertising
background image

4.3 Determinations

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

895 Professional PVC Thermomat

Connect the white silicone tubing which is connected to the reac-
tion vessel covers to the M8/olive tubing adapter of the measur-
ing vessel covers.

Screw the pieces of FEP tubing 250 mm to the M8/M6 thread
adapters of the reaction vessel covers and the air supply connec-
tors of the 895 Professional PVC Thermomat.

Insert the prepared reaction vessels into the recesses of the heat-
ing block.

Press the start button to start the data recording immediately
after inserting each single reaction vessel.

4.3.3

Cleaning the accessories

1 Cleaning the measuring vessels and accessories

Clean used measuring vessels after pouring off the measuring
solution with ethanol or 2-propanol (do not use acetone!) or in
the dishwasher.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

Pre-clean with dishwashing detergent in the case of severe con-
tamination.

Thoroughly rinse with distilled water.

Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and
thoroughly rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Remove the protective ring to facilitate cleaning of the electrodes.

2 Cleaning the reaction vessels and accessories

Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.

Clean the reaction vessel covers with acetone or 2-propanol or
in the dishwasher and rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.

NOTE

Replace the reaction vessel covers if they no longer sit tightly on
the reaction vessel or the material is brittle.