7.5 Gas flow regulation

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

895 Professional PVC Thermomat

7.5

Gas flow regulation

Volumetric flow
range

1 - 25 L/h at 25 °C and 1,013 mbar

Maximum devia-
tion from the set
range

±(0.25 L/h + 5% of the measured value)

7.6

USB interface

USB connector

USB plug type B

7.7

Power socket

Nominal voltage
range

100 - 120 V and 220 - 240 V ± 10% (autosensing)

Frequency

50 and 60 Hz (autosensing)

Power consump-
tion

450 VA

max

Fuse

Diameter 5 mm, length 20 mm
4.0 ATH (slow-acting)

7.8

Safety specifications

This instrument fulfills the following electrical safety requirements:

CE marking in accordance with the EU directives:

2006/95/EC (Low Voltage Directive, LVD)

2004/108/EC (EMC Directive, EMC)

Design and testing

According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1 protection
class I, EN/IEC 61010-2-010, EN/IEC 60529, degree of protection IP20.

Safety instructions

This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.

Advertising